Docsity
Docsity

Prepare-se para as provas
Prepare-se para as provas

Estude fácil! Tem muito documento disponível na Docsity


Ganhe pontos para baixar
Ganhe pontos para baixar

Ganhe pontos ajudando outros esrudantes ou compre um plano Premium


Guias e Dicas
Guias e Dicas

Datasheet Fluke PVA 1500, Manuais, Projetos, Pesquisas de Circuitos Elétricos

Datasheet do Fluke PVA 1500 HE.

Tipologia: Manuais, Projetos, Pesquisas

2024

Compartilhado em 12/07/2024

unica-instrumentacao
unica-instrumentacao 🇧🇷

1 / 5

Toggle sidebar

Esta página não é visível na pré-visualização

Não perca as partes importantes!

bg1
4612 . T 2012 • XP2i kPa
Page 1 of 5
ACCURACY
Standard Accuracy Version (-10 to 50° C)
0 to 20% of Range: ± (0.02% of Full Scale*)
20% to 110% of Range: ± (0.1% of Reading)
Vacuum: ± (0.25% of Full Scale**)
-S5 Accuracy Option (-10 to 50° C)
Vacuum to 110% of Range: ± (0.05% of Full Scale*)
-S2 Accuracy Option
Vacuum to 110% of Range: ± (0.02% of Full Scale*) 15 to 25° C
± (0.05% of Full Scale*) -10 to 15° C,
25 to 50° C, typical
OPERATING TEMPERATURE
Temperature Range: -10 to 50° C (14 to 122° F)
DISPLAY
Screen: 5.5 digits
Single Line Display: 16.9 mm (0.67") high
Dual Line Display: Main Display: 13.9 mm (0.55")
Secondary Display: 5.3 mm (0.21")
Display Rate: 4 readings/second (standard)
8 readings/second (PSV mode)
Includes all effects of linearity, hysteresis, repeatability,
temperature, and stability for one year.
All models indicate vacuum, but vacuum specification applies
to 100 kPa, 200 kPa, 700 kPa, and 2000 kPa models only.
Not recommended for continuous use below -100.0 kPa.
Refer to XP2i-DP data sheet for gauges that are intended for
continuous high vacuum use.
*Full Scale is 100% of the range
**Full Scale = -100.0 kPa
2000 kPa
PORT POSITION
FOR -F4 AND
-RP OPTIONS
INTERNAL PRESSUR IZED VOLUME:
295 mm3 (0.02 in 3)
SENSOR
DIAPHR AGM
SURFA CE
36.1
(1.42)
18.0
(0. 71)
43.0
(1.70)
(Ø 114 .3
4.50)
7/16-20 MP CPF PORT
19 (3/4 ")
WRENCH HEX
83.8
(3.30)
PANEL MOUNT
FLANGE
MAX PANEL
THICKNESS
11.7 (0 .4 6)
23.6
(0.93)
ADDITIONAL SENSOR LENGTH WITH OPTIONAL FITTING ADAPTERS
U.S. PATENT D612277
ALL DIMENSIONS I N MM (IN)
1/4 INCH MNPT
ADAPTER
29.1
(1.15)
1/4 INCH MBSP
ADAPTER
26.1
(1.03)
M20X1. 5
ADAPTER
32.6
(1.30 )
Non-condensing. No change in accuracy over operating
temperature range. Gauge must be zeroed to achieve rated
specification.
LCD readable in sunlight with bright backlight.
PSV mode is intended for Pressure Safety Valve testing, and
requires software to enable. (ConfigXP is a free download
from ametekcalibration.com.)
XP2i Pressure Gauge kPa 
pf3
pf4
pf5

Pré-visualização parcial do texto

Baixe Datasheet Fluke PVA 1500 e outras Manuais, Projetos, Pesquisas em PDF para Circuitos Elétricos, somente na Docsity!

  1. T 2012 • XP2i kPa

„ ACCURACY

Standard Accuracy Version (-10 to 50° C) 0 to 20% of Range: ± (0.02% of Full Scale)* 20% to 110% of Range: ± (0.1% of Reading) Vacuum: ± (0.25% of Full Scale)**

-S5 Accuracy Option (-10 to 50° C) Vacuum to 110% of Range: ± (0.05% of Full Scale)*

-S2 Accuracy Option Vacuum to 110% of Range: ± (0.02% of Full Scale)* 15 to 25° C ± (0.05% of Full Scale)* -10 to 15° C, 25 to 50° C, typical

OPERATING TEMPERATURE

Temperature Range: -10 to 50° C (14 to 122° F)

DISPLAY

Screen: 5.5 digits Single Line Display: 16.9 mm (0.67 " ) high Dual Line Display: Main Display: 13.9 mm (0.55 " ) Secondary Display: 5.3 mm (0.21 " ) Display Rate: 4 readings/second (standard) 8 readings/second (PSV mode)

Includes all effects of linearity, hysteresis, repeatability, temperature, and stability for one year. All models indicate vacuum, but vacuum specification applies to 100 kPa, 200 kPa, 700 kPa, and 2000 kPa models only. Not recommended for continuous use below -100.0 kPa. Refer to XP2i-DP data sheet for gauges that are intended for continuous high vacuum use. ***** Full Scale is 100% of the range ****** Full Scale = -100.0 kPa

2000 kPa

PORT POSITION FOR -F4 AND -RP OPTIONS

INTERNAL PRESSURIZED VOLUME: 295 mm 3 (0.02 in 3 )

SENSOR DIAPHRAGM SURFACE

(1.42)

(0.71)

(1.70)

(Ø 114. (Ø 4.50)

7/16-20 MP CPF PORT

19 (3/4") WRENCH HEX

(3.30)

PANEL MOUNT FLANGE MAX PANEL THICKNESS 11.7 (0.46)

(0.93)

ADDITIONAL SENSOR LENGTH WITH OPTIONAL FITTING ADAPTERS

U.S. PATENT D ALL DIMENSIONS IN MM (IN)

1/4 INCH MNPT ADAPTER

(1.15)

1/4 INCH MBSP ADAPTER

(1.03)

M20X1. ADAPTER

(1.30)

Non-condensing. No change in accuracy over operating temperature range. Gauge must be zeroed to achieve rated specification.

LCD readable in sunlight with bright backlight. PSV mode is intended for Pressure Safety Valve testing, and requires software to enable. (ConfigXP is a free download from ametekcalibration.com. )

  1. T 2012 • XP2i kPa

„ POWER

The XP2i is Intrinsically Safe only if powered by one of the following battery types.

ATEX/IECEx: Approved Battery Type Ta= Marking Rayovac Max Plus 815 -20 to 50° C Ex ia IIC T4 Ga Energizer E Energizer EN -20 to 50° C Ex ia IIC T3 Ga Duracell MN

CSA: Approved Battery Type Ta= Marking Rayovac Max Plus 815 -20 to 45° C Class I, Division 1, A, B, C, D T Energizer E Energizer EN91 -20 to 50° C Class I, Division 1, A, B, C, D T3C Rayovac Max Plus 815

3 x AA: 1500 hours, typical Ultra Low Power: **> 1 year, typical****

„ DATA/COMMUNICATION

Digital Interface: RS232, DB-9F

„ DATALOGGING

Capacity: 32 000 data points Storage Type: Non-volatile flash memory Minimum Interval: 1 second Maximum Interval: 18 hours Min. Event Duration: **250 mS*****

„ CERTIFICATIONS

II 1G Ex ia IIC T4/T3 Ga FTZU 12 ATEX 0048X

Ex ia IIC T4/T3 Ga IECEx FTZU 12.0009X

Exia Intrinsically Safe and Non-incendive for Hazardous Locations: Class I, Division 1, Groups A, B, C and D, Temperature Code T4 or T3C. For hazardous location product warnings, refer to the operation manual.

XP2i complies with the Electromagnetic Compatibility and the Pressure Equipment Directives.

XP2i complies with the Australian Radiocommunications (Electromagnetic Compatibility) Standard 2008.

XP2i is approved for use as a portable test instrument for Marine use and complies with Det Norsjke Veritas’ Rules for Classification of Ships, High Speed & Light Craft and Offshore Standards.

ATEX

Uses 3 alkaline AA (LR6) batteries. Use of backlight and/or RS232 reduces operating time. Optional AC power supply available. **20 minute recording interval or greater in LT5 mode. For hazardous location product warnings, refer to the operation manual.

For hazardous location product warnings, refer to the operation manual.

DataLoggerXP Firmware Update may be purchased at time of order, or an Upgrade Package may be purchased afterward. Software available for download from our website. Produces csv files, or uses Excel template files (samples included) to automatically format and graph data. *** Logging Type - Average with Peaks.

  1. T 2012 • XP2i kP a

P/N Model (^) DisplayDual Accuracy AdapterType ConnectionLocation Data logging Pump System Carrying Case~ XP2I* Standard: No.. (omit) 0.1% of reading.. (omit) NPT..... .(omit) Bottom.... (omit) No...... .(omit) No Pump.... (omit) Optional: Yes.... -DD 0.02% of FS........ -S2 G 1/4 B.... -BSP Back......... -RP Yes......... -DL System A.... -AXX Aluminum... (omit) 0.05% of FS........ -S5 M20x1.5.. -M20 Panel Mount.. -F4 System A.... -AHX Waterproof...... -W System B.... -BXX System B.... -BHX System C.... -CXX System C.... -CHX System D.... -DOX System D.... -DWX System E.... -EOX System F.... -FOV System F.... -FWV System G.... -GOX System G.... -GWX System H... -HOX

~ The Waterproof Case is an option for Systems A, B, and C only. The Waterproof Case is the only option for Systems G and H.

AMETEK offers a variety of solutions for pressure genera- tion and measurement. Our line of products for pressure generation includes everything from small pneumatic hand pumps to a precision, hydraulic pressure comparator capable of generating up to 15 000 psi / 1000 bar / 100 MPa.

All of our pumps may be ordered as part of a Pump System, complete with an XP2i and delivered in a sturdy carrying case with custom insert.

  • Refer to the following page for a more detailed description of each pump system.

„ RANGE & RESOLUTION TABLE

Display Resolution P/N Range(kPa) pressureOver- kPa MPa bar mbar

100KPA 100 6.5 x 0.01 0.00001 0.0001 0. 200KPA 200 3.0 x 0.01 0.00001 0.0001 0. 700KPA 700 2.0 x 0.01 0.00001 0.0001 0. 2KKPA 2000 2.0 x 0.1 0.0001 0.001 1 3KKPA 3000 2.0 x 0.1 0.0001 0.001 1 7KKPA 7000 2.0 x 0.1 0.0001 0. 14KKPA 14 000 2.0 x 1 0.001 0. 20KKPA 20 000 1.5 x 1 0.001 0. 30KKPA 30 000 1.5 x 1 0.001 0. 70KKPA 70 000 1.5 x 1 0.001 0. 100KKPA 100 000 1.3 x 1 0.001 0.

„ ORDERING INFORMATION

*KPA versions available in USA direct from factory only.

X Ordering a Pump System Only

Any pump system, carrying case, and connection fittings for an XP2i pressure gauge may be ordered separately from the gauge. Enter XP2I-NONE followed by the Pump System part number and the Carrying Case option code. SAMPLE PART NUMBER XP2I-NONE-BHX-W................. System B pump system with a waterproof carrying case.

CPF Adapter Fitting is not included.

SAMPLE PART NUMBERS

200KPAXP2I..................... 200 kPa standard gauge, with a 1/4" NPT pressure fitting. 30KKPAXP2I-S2-BSP-DL........ 30 000 kPa gauge with 0.02% of FS accuracy, a 1/4" BSP pressure fitting, and Data logging. 70KKPAXP2I-DD-GWX-W........ 70 000 kPa Dual Display gauge with a System G pump system and a waterproof carrying case.

Page 5 of 5

ametekcalibration.com 4612. T 2012 • XP2i kP a © 2020 Crystal Engineering Corporation 708 Fiero Lane, Suite 9, San Luis Obispo, California 93401-

„ PUMP SYSTEMS OVERVIEW

Pump System

Case Options Part Number Pressure Range Pneumatic Hydraulic Hand Pump Bench Top Included Pump Aluminum Waterproof (Pelican Case)

System A

AXX 0 to 30psi / 2 bar „ „ T-960-CPF „ „

AHX 0 to 580 psi / 40 bar „ „ T-970-CPF „ „

System B

BXX -25 inHg to 30 psi / -0.85 to 2 bar „ „ T-965-CPF „ „

BHX -27 inHg to 580 psi / -0.91 to 40 bar „ „ T-975-CPF „ „

System C

CXX 0 to 3000 psi / 200 bar „ (Oil) „ T-620-CPF „ „

CHX 0 to 5000 psi / 350 bar „ (Oil) „ T-620H-CPF „ „

System D

DOX 0 to 5000 psi / 350 bar „ (Oil) „ P-018-CPF „

DWX 0 to 5000 psi / 350 bar „ (Water) „ „

System E EOX 0 to 10 000 psi / 700 bar „ (Oil) „

P014-CPF

System F

FOV 0 to 15 000 psi / 1000 bar „ (Oil) „ T-1-CPF „

FWV 0 to 15 000 psi / 1000 bar „ (Water) „ „

System G

GOX 0 to 15 000 psi / 1000 bar „ (Oil) „ GaugeCalHP „

GWX 0 to 15 000 psi / 1000 bar „ (Water) „ „

System H HOX

-27 inHg to 580 psi / -0.91 to 40 bar „ „ T-975-CPF „

0 to 5000 psi / 350 bar „ (Oil) „ T-620H-CPF „

(or)

(or)

(or)

(and)